9 Diagnostics and service
VEGAFLEX 83 • Profibus PA
44223-EN-140430
1
Fig. 47: Exchanging the measuring rod
1 Seal ring
Information:
Please maintain the specified torque so that the max. tensile strength
remains.
6. Enter new probe length and if necessary the new probe type and
then carry out a fresh adjustment (see "
Setup procedure, Carry
-
ing out min. adjustment - Carrying out max. adjustment
").
9.7 Software update
The following components are required to update the instrument
software:
•
Instrument
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current instrument software as file
You can find the current instrument software as well as detailed
information on the procedure under "www.vega.com/downloads" and
"
Software
".
Caution:
Instruments with approvals can be bound to certain software versions.
Therefore make sure that the approval is still effective after a software
update is carried out.