3.2 Principle of operation
VEGACAL 65 is a level sensor for use in all areas of industry.
The partly insulated probe is ideal for measurement of bulk solids and
can be also used in non-conductive liquids such as for example oil.
The proven mechanical construction offers high functional safety.
Probe, measured product and vessel wall form an electrical capacitor.
The capacitance is influenced by three main factors.
1
2
3
Fig. 3: Functional principle - Plate capacitor
1 Distance between the electrode surfaces
2 Size of the electrode surfaces
3 Type of dielectric between the electrodes
The probe and the vessel wall are the capacitor plates. The measured
product is the dielectric. Due to the higher dielectric constant of the
product compared to air, the capacitance increases as the probe is
gradually covered.
The capacitance as well as the resistance change are converted by
the electronics module into a level-proportional signal.
Power is supplied via the H1 Fieldbus. A two-wire cable according to
Fieldbus specification serves as carrier of both power and digital data
for multiple sensors. This cable can be operated in two versions:
•
via an H1 interface card in the control system and additional power
supply
•
via a Linking device with HSE (High speed Ethernet) and addi-
tional power supply according to IEC 61158-2
The DD (Device Descriptions) and CFF (capability files) necessary
for planning and configuration of your FF (Foundation Fieldbus) com
-
munication network are available in the download area of the VEGA
homepage www.vega.com
. The appropriate certificates are also
available there. A CD with the appropriate files and certificates can be
ordered via e-mail under [email protected] or by phone from one of
the VEGA agencies under the order number "DRIVER.S".
The backlight of the display and adjustment module is powered by the
sensor. Prerequisite is a certain level of operating voltage.
Application area
Functional principle
Power supply and bus
communication
DD/CFF