9 Diagnostics and servicing
SOLITRAC 31 • Profibus PA
43834-EN-211203
This status message is inactive by default.
Code
Text message
Cause
Rectification
F008
Error multi sensor commu-
nication
Additional sensors not switched on
EMC influences
No other sensor available
Check wiring between the sensors
Connect the sensors correctly and make
them ready for operation
F013
Sensor signals a fault
Error on the current input/digital input
No valid measured value
Connected instruments without function
Check current input
Check connected instruments (Second-
ary instrument)
F016
Adjustment data exchanged
Values of the min. and max. adjustment
exchanged
Correct adjustment data
F017
Adjustment span too small
The values of the min. and max. adjust-
ment are too close together
Correct adjustment data
F025
Invalid linearization table
Wrong or empty linearization table
(1074, 1075, 1080, 1100, 1106)
Wrong value in the linearization table
(1143, 1144)
Create linearization table
Correct linearization table
F030
Process value out of limits
Process values are not within the adjust-
ed measuring range
Repeat adjustment
F034
EPROM hardware error
Electronics defective
Restart instrument
Exchanging the electronics
F035
EPROM data error
Error in the internal instrument commu-
nication
Carry out a reset
Exchanging the electronics
F036
Faulty program memory
Error during software update
Repeat software update
Exchanging the electronics
F037
RAM hardware error
Error in RAM
Restart instrument
Exchanging the electronics
F038
Secondary signals failure
Connection cable to the Secondary in-
strument interrupted
Instrument not defined as Secondary in-
strument
One of the Secondary instruments sig-
nals a failure
Check the connection cable to the Sec-
ondary instrument
Define instrument as Secondary
Check Secondary instruments
F040
Hardware error
Instrument defective (1092, 1126)
Temperature outside the specification
(1091)
Restart instrument
Exchanging the electronics
Cool the instrument or protect it with iso-
lation material against heat/cold
F041
Photomultiplier error
Error in the measured value recording
Restart instrument
Exchanging the electronics
F052
Faulty configuration
Invalid parameter adjustment
Carry out a reset
Failure