EMI Shielding, Airflow, Air Filter and Fan Trays
90
3.
Avoid Electric Overload
: To avoid electric shock, fire hazard, or damage to the
instrument, do not apply a higher voltage.
a. The DC Power Supply is designed to be powered from -48 VDC, 420
Watts.
b. A Power Load meter shows the current power load from 0% to 100%.
4.
Ground the Instrument
: To avoid electric shock using a DC power supply,
ensure that the chassis is properly grounded to earth ground before making input
power connections to the device. A ground connector is provided on the back
side of the chassis for the use of a grounding lug or ground wires with a single U-
shaped connector or Ring type connector.
EMI Shielding, Airflow, Air Filter and Fan Trays
Warning!
Never turn the MPA chassis Power on with any of the bays/slots
unpopulated or uncovered!
EMI Shielding and Airflow Warning
Before turning the chassis Power On, verify that the Fan Trays, the MCH controller
module, and all of the AMC cards are properly inserted, and verify that any empty bays
have blank filler plate covers inserted.
MPA_e_manual_D07-00-129P_RevA00