Protocol Tabs
361
(Specifically for SONET testing)
Performance Monitoring provides greater detail about incoming Section, Line, Path, VT
and BIT errors.
The following options are available:
•
•
•
•
(only available when in a VT mapping)
•
Perf Monitor - Sect
(Specifically for SONET testing)
The Section screen provides a detailed breakdown for Section errors detected and
logged by the SONET protocol processor.
The following statistics are reported:
Section Near End B1 Errors
•
CV
-
Coding Violations
. Reports an error count of incoming B1 errors.
•
ES
-
Errored Seconds
. Reports the duration of the incoming B1 errors in
seconds and the percentage of errored seconds compared to the duration of
error free seconds.
•
EFS
-
Error Free Seconds
. Reports the duration of the incoming B1 error free
seconds and the percentage of error free seconds compared to the duration of
errored seconds.
•
SES
-
Severely Errored Seconds
. Reports the duration of Severely Errored
Seconds (SES) on the incoming SONET signal. An SES is reported when the
number of B1 errors per second equals or exceeds the
Section SES
Threshold
(see table below), or when a Severely Errored Frame (SEF) alarm,
Loss of Frame (LOF) alarm, or Loss of Signal (LOS) alarm is present.
Interface
Section SES threshold
MPA_e_manual_D07-00-129P_RevA00
Summary of Contents for MPA
Page 2: ...MPA_e_manual_D07 00 129P_RevA00...
Page 10: ...MPA_e_manual_D07 00 129P_RevA00...
Page 82: ...MPA_e_manual_D07 00 129P_RevA00...
Page 110: ...MPA_e_manual_D07 00 129P_RevA00...
Page 134: ...MLD Tab 134 100G RS FEC Ethernet 400G RS FEC Ethernet MPA_e_manual_D07 00 129P_RevA00...
Page 255: ...Protocol Tabs 255 MPA_e_manual_D07 00 129P_RevA00...
Page 256: ...OTN Tab 256 MPA_e_manual_D07 00 129P_RevA00...
Page 748: ...MPA_e_manual_D07 00 129P_RevA00...
Page 796: ...MPA_e_manual_D07 00 129P_RevA00...