Figure 5-63. Application of O-Rings
12. With propanol and Tex-wipe, clean the bottom surface of the Viewport Assembly, consisting of
Viewport Cover and Viewport Adapter.
Figure 5-64. Viewport Adapter Cleaning
13. Take note of the direction of the two glands on the side of the Viewport Adapter and position them so
that they face the same side as the front of the Flow Flange (marked