Stage System
Mounting of Samples
70
Dimension 3100 Manual
Rev. D
5.1
Mounting of Samples
There are two methods widely used for mounting samples: vacuum chucks and magnetic pucks.
Regardless of the method used, verify that samples are mounted flat and parallel to the stage. This
is especially important for larger samples inspected over more than one site. Grossly tilted samples
may require raising the head higher whenever the sample is indexed, increasing cycle time and the
risk of probe-sample collision. Similarly, large numbers of identical samples should be mounted
the same way whenever possible to allow use of the same settings between samples.
5.1.1 Vacuum Chucks
Stages are equipped with vacuum chucks which are often employed for securing samples.
5.1.2 Magnetic Pucks
Samples may also be secured to the stage using magnetic pucks. This system allows for easy
mounting and removal of small samples. For detailed mounting instructions, see
.
CAUTION:
Operators should be cautious when handling larger samples with
the vacuum chuck—an 8-inch wafer held using 5 psi vacuum
sustains a loading of 250 lbs.
CAUTION:
If debris is trapped between the stage and wafer, the wafer may
become scratched or broken. Keep the stage clean at all times using
isopropyl alcohol. Dispose of wipes in an appropriately labelled
solvent-contaminated waste container. When cleaning dust from
the stage area, use a vacuum cleaner with a soft brush. DO NOT
clean stages using compressed air.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......