Force Imaging
Force Modulation
Rev. D
Dimension 3100 Manual
237
Figure 13.6c
Auto Tune Controls Panel
d. Set the
Drive frequency
to
15kHz
and the
Sweep width
to
30kHz
.
e.
Set the
Drive amplitude
at
1V
to start. You can readjust this value later.
Note:
A series of peaks display on the
Frequency Sweep
plot. A typical
Frequency
Sweep
plot is shown in
Figure 13.6d
Typical Frequency Sweep Plot
Note:
The large drive amplitude is necessary because peaks are smaller than normally
seen during TappingMode operation. This is due to the cantilever not at
resonance; therefore, its motion is mostly vertical. Vertical motion is not
amplified by the beam deflection detection technique which is sensitive
primarily to changes in cantilever angle.
019
Peaks due to bimorph
Peaks due to cantilever