Force Imaging
Force Calibration (Contact Mode AFM)
222
Dimension 3100 Manual
Rev. D
13.4.2 Helpful Suggestions
To minimize or calculate the contact force between the tip and sample, it is important to obtain a
good force curve; however, there will be situations where you will not obtain a good force curve no
matter how much you adjust the
Ramp size
,
Z scan start
and
Setpoint
. The following sections
discuss a few such situations. Suggestions on working in
Force Calibration
mode are included.
False Engagement
illustrates a force curve resulting from a falsely engaged tip. The photodiode receives
light reflected off the sample, causing an increase in the deflection signal until the signal equals the
setpoint and the system “engages” (even though the tip is not on the surface). Interference in the
reflected light causes the hump-shaped waveform.
The easiest way to correct a false engagement is to withdraw the tip, adjust the photodetector
positioner to make the top/bottom differential voltage more negative, then re-engage the tip. This
compensates for the slight slope in
Figure 13.4c
False Engagement (G Scanner)
Motor Control
Motor Control
>
Tip Up
and
Tip Down
buttons provide coarse adjustment of
Z center
voltage.
With these buttons the SPM head moves vertically. The feedback loop causes the Z piezo to adjust
to compensate for the head movement. If you use
Tip Up
, the Z piezo extends. If you use
Tip
Down
, the Z piezo retracts.
Note:
The
Step Motor
function is generally used only when the scanning range of the
Z piezo is exceeded or when it is necessary to position a force measurement in
the center of the scanner range. Because the Z-axis leadscrew has some
backlash, it may be necessary to rotate the leadscrew several turns by clicking
on the
Tip Up
or
Tip Down
buttons before obtaining movement.
Tip
Deflection
0.48 V/div
Z Position - 9.27 V/div
Setpoint
Retracting
Extending
Possible False Engage Points
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......