TappingMode AFM
Advanced TappingMode AFM Operation
Rev. D
Dimension 3100 Manual
145
Figure 9.4c
Scope Trace with Correct Scan Rate
9.4.3 Decreasing the Cantilever Drive Frequency
The
Drive frequency
selected to oscillate the cantilever plays an important role in the performance
of the microscope while in TappingMode. As a first step it is important to determine the resonant
frequency of the cantilever, but the
Drive frequency
could be further tuned to improve scanning
performance.
The microscope produces better data in TappingMode when the
Drive frequency
is set lower than
the resonant peak of the cantilever. The
Drive frequency
is set such that it coincides with a 1-10
percent decrease in the oscillation amplitude by setting
Peak offset
in the
Auto Tune
controls to the
desired percent.
shows a suggested operating region. This is a suggestion based on our
observations; users are encouraged to experiment with the microscope and decide what produces
the best results.
Figure 9.4d
Suggested Range of Drive Frequencies
Z Range
50.00 nm/div
Scan Size - 2.50 µm/div
Trace
Retrace
Cantilever
Response
12.54 nm/div
Center Frequency - 338.8723 KHz
5.0 KHz/div
Setpoint
Drive Frequency
should
be within this range.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......