TappingMode AFM
Basic TappingMode AFM Operation
140
Dimension 3100 Manual
Rev. D
9.2.9 Set Initial Scan Parameters
Scan Controls Panel
In the
Scan Controls
panel, set the following initial scan parameters (see
1. Set the
Scan Rate
to
2 Hz
.
2. Set the
Scan Size
to
1µm
.
3. Set the
Scan Angle
to
0
.
4. Set
X
and
Y Offsets
to
0
.
Figure 9.2f
Suggested Scan Controls Settings
Other Controls Panel
Verify the
Z Limit
is at the maximum value.
Figure 9.2g
Suggested Other Controls Settings
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......