TappingMode AFM
Basic TappingMode AFM Operation
136
Dimension 3100 Manual
Rev. D
Automatic Tuning
For most purposes, the
Auto Tune
function will suffice.
•
Verify the starting parameters. The nominal resonant frequency of a probe can be found
on the probe box label. Ensure the start and end frequencies correlate with this value.
•
Click
Auto Tune
. The computer and controller begin automatic tuning (see
Figure 9.2a
Auto Tune Control Panel
Manual Cantilever Tuning
With Force Modulation or Fluid Tapping applications, it may be useful to tune the cantilever
manually.
Note:
More than one type of cantilever exists. Cantilevers can have different
dimensions and different resonant frequencies. Certain parameter values,
particularly the center frequency and the sweep width used in the following
example, apply to a particular cantilever type. In the following example,
nominal parameter values will vary depending on the actual cantilever used.
1. In the
Cantilever Tune
panel, set the
Drive frequency
parameter to a value near the center
of the resonant frequencies range specified for the wafer. For example, if the frequency range
is specified as 240—420 KHz, select a drive frequency of
330 KHz
.
2. Set the
Drive amplitude
to
200 mV
.
3. Set the
Sweep Width
to the same value as the
Center Frequency
.
4. The
Sweep Width
must be large enough to cover the frequency range specified for the
wafer.
5. Zero the
Setpoint
.
6. Set the
Amplitude limit
to
2.5V
.
7. Center the peak on the frequency sweep plot shown on the display monitor using the
Zoom
In
and
Offset
commands after identifying the maximum amplitude peak with the lowest
frequency in the frequency response plot.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......