Head, Probe, & Sample Preparation
Basic AFM Operation
102
Dimension 3100 Manual
Rev. D
7.2
Basic AFM Operation
7.2.1 Select the Microscope
1. Set the microscope configuration by selecting the
Microscope Select
panel from the
Di
pop-
down menu (see
2. Click
OK
to close the
Microscope
>
Select
dialog box and initiate the Realtime software.
Figure 7.2a
Microscope Select Prompt
7.2.2 Select Mode of Operation
1. Select
Microscope
>
Profile
.
2. Select the desired mode of operation (Contact Mode, TappingMode, etc.) and click
OK
.
7.2.3 Prepare the Cantilever Holder
Before loading the substrate onto the cantilever holder, place the cantilever holder on the cantilever
holder stand (see
). The cantilever holder stand is a 2.5-inch diameter black anodized
aluminum cylinder that ships with the Dimension 3100 SPM for easy installation of the substrate
onto the cantilever holder.
The cantilever holder stand has three stations for different tip holders: 1) standard AFM, 2) fluid
imaging AFM and 3) STM. A small, raised block in the center of the gold connector identifies the
standard AFM load station.
To prepare the cantilever holder, with the AFM cantilever holder's large spring clip face up, mate
the standard AFM cantilever holder sockets to the pins of the AFM cantilever holder stand (see
).
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......