Single Scan Operation
4-15
SETTING THE ZERO POINT
You may select any point on the profile trace as the zero point. The zero point is the point of reference from
which all measurements are taken. Software leveling sets both the
R
and
M
cursor intercepts at zero.
However, when the
Zero
function is activated, it sets the zero point only at the R cursor intercept.
To set the zero point:
1
Position the
R
(reference) cursor at the desired zero location.
2
Select
Plot
>
Zero
(or press
F8
on the keyboard) to automatically replot the profile trace and establish the
zero point at the
R
cursor intercept (see
).
Figure 4-15: Setting the Zero Point
MAKING A DELTA AVERAGE STEP HEIGHT
MEASUREMENT
Once you run the scan routine and leveled the profile, you can obtain an accurate vertical
measurement of the calibration standard using the Delta Average Step Height analytical function.
Analytical functions are calculated using the
R
and
M
cursors.
1
In the
Data Plo
t window, click the
Analytical Functions
icon
or select
Plot
>
Analytical
Functions
. Alternatively, right-click on the plot and select
Analytical Functions....Insert
. The
Analytical
Functions
dialog box appears
.
Zero Point
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......