Scan Routine Parameters
7-15
NOTE –
The
End Force
that will be applied by the stylus is calculated and displayed.
Figure 7-16: Scan Parameter Dialog Box: Deflection Parameters (Optional)
Example:
•
Scan Duration
is set at 13 seconds.
•
Stylus Force
is set at 3 mg (on the
Nominal Parameters
tab).
•
Interval
is set at 1000 msec (1 second).
•
Increment
is set at 0.05 mg.
•
End Force
is calculated and displayed as 3.65 mg (13 seconds divided by 1 second, multiplied
by 0.05 mg = 0.65 mg to be added in 0.05-mg increments to 3 mg).
DISPLAY PARAMETERS
In its
Display Parameters
section, the
Scan Routines
window contains parameters that allow
automatic manipulation of the graphic display of the profile trace. The display parameters are
described below.