Scan Routine Parameters
7-11
Figure 7-11: Scan Parameters Dialog Box: Measurement Range
Profile
The
Profile
setting scales the measurement range according to the profile selected. Three different
profiles are available for a variety of sample surface characteristics (see
Figure 7-12: Sample Surface Profiles
•
Valleys:
Provides 90% of the measurement range below the zero horizontal grid line. Used primarily for
measuring etch depths.
•
Hills and Valleys:
Provides 50% of the measurement range above the zero horizontal grid line and 50
percent below. Used in most applications, especially if the surface characteristics of the sample are not
well known, or if the sample is out of level.
•
Hills
: Provides 90% of the measurement range above the horizontal grid line. Used primarily for
measuring step heights.
1
Click any parameter in the
Scan Parameters
section of the
Scan Routines
window (see
display the
Nominal Parameters
tab of the
Scan Parameters
dialog box.