Electric Force (EFM) Imaging
Electric Field Gradient Detection—Preparation
244
MultiMode SPM Instruction Manual
Rev. B
14.3 Electric Field Gradient Detection—Preparation
This section explains how to conduct electric
fi
eld gradient imaging by applying a voltage to the tip
or sample to generate electric
fi
elds.
Note:
If the sample being imaged has a permanent electric
fi
eld which does not
require the external application of voltage, the steps below are not required and
you can proceed to
.
Please consider the following before proceeding:
•
Before attempting to recon
fi
gure the jumpers, carefully read the following Jumper
Con
fi
guration sections.
•
A modi
fi
ed cantilever is required to bias the tip in EFM. Contact Veeco tech support for
more information.
When it is necessary to apply voltage to the tip or sample, minor changes must be made to the
jumpers in the microscope’s baseplate and the toggle switches on the Basic Extender Module (if
equipped). Original jumper con
fi
gurations and jumper changes are dependent on the microscope
being used and the measurements desired.
provides jumper con
fi
guration
instructions for basic microscope models operating
without
the Basic Extender Module in four
sections:
•
“Voltage Applied to the Tip” on page 246
•
“Voltage Applied to the Sample” on page 248
•
“External Voltage Source Applied to the Tip” on page 249
•
“External Voltage Source Applied to the Sample” on page 250
provides jumper con
fi
guration instructions for basic microscope models operating
with
the Basic Extender Module in four sections:
•
“Voltage Applied to the Tip” on page 252
•
“Voltage Applied to the Sample” on page 253
•
“External Voltage Source Applied to the Tip” on page 254
•
“External Voltage Source Applied to the Sample” on page 255
The location and orientation of the jumpers in the baseplate of the MultiMode is shown below in
. To change the jumpers, it should not be necessary to remove the baseplate; they can
easily be changed through the rectangular opening in the bottom of the baseplate using a pair of
non-conducting tweezers. For non-EFM applications and surface potential operation, jumpers are
usually left in their original positions or returned to their original positions.