
Force Imaging
Force Calibration Control Panels and Menus
Rev. B
MultiMode SPM Instruction Manual
183
11.3 Force Calibration Control Panels and Menus
The
Force
Calibration
Control
window (see
) manipulates the microscope in
Force
Calibration
mode. The parameters control the rate, start position and amplitude of the triangle
wave applied to the Z piezo. You can also adjust the
Setpoint
value of the cantilever de
fl
ection
voltage used in the feedback loop during imaging. The
Capture
button stores the force curve for
Of
fl
ine
viewing. Some parameters in
fl
uence the operation of the microscope during imaging; most
of the menu parameters affect only
Force Calibration
mode.
Figure 11.3a
Advanced Force Calibration Control Window (Contact Mode AFM)
There are 2 methods to enter Force Calibration Mode:
•
View
>
Force Mode
>
Calibrate
will open a simpli
fi
ed subset of the parameters and
commands available.
•
View
>
Force Mode
>
Advanced
includes the full set of available parameters and
commands.
Additionally:
•
View
>
Force Mode
>
Step
enters force mode with the parameter
start mode
=
step
and the piezo fully retracted.
•
View
>
Force Mode
>
Volume
enters Force Volume Mode.
Items in the
Force Calibration
Control
Window are discussed individually below.
038