Vectronics PM-30 Quick Manual Download Page 3

2. The Peak Reading feature may be switch on/off by using the 
pushbutton switch labeled POWER PEAK/AVG. Peak reading is 
activated when the switch is locked in.

3. SWR is indicated at the intersection of the two meter pointers. 
While transmitting, read the SWR from the red SWR line nearest the 
intersecting point.

CALIBRATION

To null the directional coupler, apply power through the RF INPUT 
connector, terminating the output in 50Ω. Adjust the power level for 
50-100 Watts on the 300 W range. Looking from the front of the 
meter, adjust the left-hand trim capacitor on the directional coupler 
board for a null or minimum meter deflection. Reverse the input and 
output cables and again apply power. Adjust the right-hand trim 
capacitor for a null on the other meter.

Using a calibrated source, adjust the meter with the internal trimpots 
as label in the following diagram: