6 Data Interface
15 / 41
6.1.3
Output Data Specification
The image data is output as 12 bit values. A bit assignment for 12 bit grey values according to the
Camera Link specification is used.
Camera Link Port
(1 x 12 Bit)
A0 D0
(LSB)
A1 D1
A2 D2
A3 D3
A4 D4
A5 D5
A6 D6
A7 D7
B0 D8
B1 D9
B2 D10
B3 D11
(MSB)
B4
B5
B6
B7
C0
C1
C2
C3
C4
C5
C6
C7
6.2
Conversion of image data to temperature data
Due to the internal LUT, the pixel information can be converted to a corresponding temperature value.
For the IRC-320 the following relation can be used:
o
DN
r
T
+
=
*
with:
: 12-bit Digital value (Pixel data)
DN
T
: Temperature value [°C]
r
: Resolution (for the IRC-320 standard version: 0.03°C)
: offset for DN=0 (for the IRC-320 standard version: -30°C)
o
The
following diagram shows the result.
VDS Vosskühler GmbH