Installation manual
17
1.13. System Characteristics
1.13.1. Mechanical Characteristics
z
Source to Image distance [Focal spot to Sensor]
603.8mm
z
Vertical column movement
Max. 700 mm
z
Weight
78
kg
z
Length x Width x Height
1150mmx950 x2200 mm
z
Type of installation
Base Stand
1.13.2. Electrical Characteristics
z
Power supply voltage
AC 110/230V ± 10%
z
Frequency
50/60
Hz
z
Power
rating
2.0KVA
1.13.3. Environmental Characteristics
z
Operating temperature
10 ~ 30
℃
z
Operating relative humidity
30 ~ 75%
z
Operating atmospheric pressure
700 ~ 1060 hPa
z
Transport and storage temperature
-20 ~ 70
℃
z
Transport and storage relative humidity
< 90% non-condensing
z
Transport and storage atmospheric pressure
500 ~ 1060 hPa