8. Acquiring CEPH Images (Optional)
Green Smart User Manual 89
EN
G
LISH
8.3
Patient Positioning
Have patience (especially pregnant women and children) wear a
lead apron to protect themselves from residual radiation.
Be careful not to shine the laser beam directly into the person's
eyes. Doing so may result in vision loss.
Ensure that the Nasal Positioner is left unfolded, before adjusting the
Ear Rods in the proper direction.
Correct posture reduces the shadow cast by the patient's cervical
spine and allows clear image acquisition.
Metal implants or bridges may reduce the quality of the images.
Getting prepared
1.
Let the patient remove all the metal objects (glasses, earrings, hair pins, braces, false
teeth, etc.). Metal objects may induce ghost images and lower the image quality.
2.
Have the patient wear a lead apron to protect themselves from residual radiation.
3.
Use the
Column UP/DOWN
button or switch option to adjust the equipment to match the
height of the patient.
< Control Panel – Membrane type>
<Control Panel – LCD type>
Summary of Contents for GREEN SMART
Page 1: ...User Manual Version 1 23 English...
Page 2: ......
Page 3: ......
Page 4: ......
Page 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Page 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Page 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Page 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Page 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Page 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Page 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Page 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Page 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Page 200: ......