Page
139
of
244
4.4.
PANO Collimator Alignment
4.4.1.
PANO Collimator Alignment
1. Run the C:\VCaptureSW\Acquisition\PANO\WidePANO\VACAP.exe file.
2. Send [SPM_PANO], [SPM_PSON], [SPM_FIPM_0004], [SPM_FISS_0001],
[SPM_FISM_0001], [SPM_COLM_0007] commands to the command window to set the
PANO sensor and collimator.
3. Send [SPM_HV___0740], [SPM_HA___1200], and [SPM_SSB_] commands in the VACAP’s
Command window, and acquire the image.
4. Run EzEval.exe file, choose EzEvaluation for the tap menu on the top left and choose
collimator viewer from the tap menu on the top right.
5. Set the threshold value concerning the following standard, and check 0010.raw to check if it
satisfies the collimator alignment standard below.
6. If the Left and Right do not meet the standards, adjust the [SPM_CCST_XXXX] command.
To check the current value, enter [SPM_CCS?] to the command window.
7. Top/Bottom should be 1~10 pixels. If it does not meet the standard, check the collimator
position and move it.
※
Collimator Alignment standard value
Collimator
Min pixels
Max pixels
Threshold
70%
Left
2
2
Right
1
2
Top
1
10
Lower
1
10
If values are changed after the collimator alignment, conduct the collimator
speed calibration. (See
4.5
)
The
threshold
70%
area is marked in red.
Set Threshold to 70 and
click the Apply button.
Left
Right
Choose the 0010raw file.
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...