Installation, Operating & Maintenance Instructions 

Series 203 + 204, DN 63 - 200 (I.D. 2½ - 8")

 

 

 

 

VAT Vakuumventile AG, CH-9469 Haag, Switzerland 
Tel +41 81 771 61 61   Fax +41 81 771 48 30   [email protected]   www.vatvalve.com 

229793EE

 

2006-03-17 

5/11 

 

2 Installation 

 

2.1  Installation into the system 

 

 

Use standard ISO centering rings only for installing ISO flanges! 

The centering rings must fit well into the centering groove. Any clearance is prohibited. Too thick or 
improperly centered rings can cause the plate to touch the centering ring if the plate is in open 
position. This would damage the plate seal or interfere with the plate movement! 

 

2.2 Connections 

 
2.2.1  Compressed air connection 

 

 

Compressed air may only be connected by an authorized expert! 
 
Connect compressed air only when   -  valve is installed into the system 
   

-  moving parts cannot be touched 

 Compressed 

air 

connection:  

 

 Internal thread M5 (

1

/

8

" optional) with valves without solenoid 

 (open/close) 

 Internal thread 

1

/

8

" with valves with solenoid 

  (IN at the solenoid) 

Compressed air pressure (min. - max. overpressure): 4 - 7 bar / 60 - 100 psi 

Use clean, dry or slightly oiled air only! 

 
2.2.2 Electrical 

connection 

 

 

Electrical power may only be connected by an authorized expert! 
 
Do not touch parts under voltage! 

 

 

Connect electrical power only when 
-  valve is installed into the system 
-  moving parts cannot be touched 
 
Verify that the control voltage is in accordance with the 
voltage stated on the solenoid!  
See (4) on drawing page 10 
 
Plug for position indicator and solenoid are supplied with 
the valve. 

MV

2

1

power

2

1

 

MV = coil of solenoid 
 
Plug (5) on drawing page 10 

 

 

Connection of position indicator plug: 

 

 

 

 

 unisolate 

connection 

strain 

relief 

 

 

1 press to open the clamp 

 

 

 

2 insert the cable to the plug 

 

5 – 6 mm 

Summary of Contents for 203 Series

Page 1: ...lar made in Switzerland Fabrication No patented 20 A Fabrication number Explanation of symbols Read declaration carefully before you start any other action Keep body parts and objects away from the va...

Page 2: ...lms or any other reproduction processes nor may it be manipulated with electronic systems duplicated or distributed without written permission from VAT Offenders are liable to pay damages The original...

Page 3: ...r connection 5 2 2 2 Electrical connection 5 3 Operation 6 3 1 Normal operation 6 3 2 Operation under increased temperature 6 3 3 Behavior in case of compressed air pressure drop 6 3 4 Behavior in cas...

Page 4: ...n permission of VAT 1 1 Technical data Leak rate body valve seat 1 10 9 mbar ls 1 Pressure range 1 10 8 mbar to 2 bar abs Differential pressure on the plate 1 2 bar in opening direction 1 6 bar in clo...

Page 5: ...ly when valve is installed into the system moving parts cannot be touched Compressed air connection Internal thread M5 1 8 optional with valves without solenoid open close Internal thread 1 8 with val...

Page 6: ...mains open in any intermediate position plate movement is undefined 3 4 Behavior in case of power failure Valve Reaction of valve supplied with solenoid closes without solenoid dependent on pneumatics...

Page 7: ...rings and or air cushions in cylinders can be potential hazards Keep fingers and objects away from the valve opening Products returned to VAT must be free of harmful substances such as e g toxical cau...

Page 8: ...t screw 3 completely but do not fasten it Attention Between plate 2 and shaft the adjustment washers which may have been dismounted when the plate was dismounted have to be reinstalled Close valve Kee...

Page 9: ...8 Connect plug to position indicator 7 Switch on electrical power 5 4 Replacement of bonnet seal feedthrough O ring 9 The item numbers in brackets refer to the drawing on page 10 Valve must be remove...

Page 10: ...ting Maintenance Instructions Series 203 204 DN 63 200 I D 2 8 VAT Vakuumventile AG CH 9469 Haag Switzerland Tel 41 81 771 61 61 Fax 41 81 771 48 30 CH vatvalve com www vatvalve com 229793EE 2006 03 1...

Page 11: ...uct or component that has been repaired or altered by anyone other than an authorized VAT representative or that has been subject to improper installation or abuse misuse negligence or accident VAT sh...

Reviews: