Chapter 4. Solids Cabinet Preparation and Installation
26
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
WARNING:
Remember that during the installation the magnet is up to field.
Equipment and tools used by the moving crew must be evaluated for
safety.
Install a plastic chain (or chain of other nonmagnetic material)
at about 42 in. (107 cm) from the magnet body. This will serve as a
reminder to prevent magnetic materials (e.g., consoles) from being
pulled into the magnet body.
4.
Verify that there is enough room to accommodate the solids cabinet on the floor plan
of the laboratory by following the guidelines below:
a.
Measure the distance between the edge of the
UNITY
INOVA console and the
centerline of the magnet.
If this distance is greater than the distance indicated in
Figure 3. Solids Cabinet Open Front View
Status panel
Sorensen VT
Low band 1 kW
amplifier
Low band
power supply
High band 1 kW
amplifier
High band EHT
power supply
controller (optional)
Table 3. Minimum Distances of
UNITY
INOVA Console from Magnet Centerline
System (MHz)
Distance
300/51
170 cm (67 in.)
300/89
185 cm (73 in.)
400/54
185 cm (73 in.)
400/89
185 cm (73 in.)
Summary of Contents for UNITY INOVA NMR
Page 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...