TECHNICAL INFORMATION

5

87-900-939-01 (D)

Serial Communication Port

Communication serial  port  connections  and  mini-  mum
connection  configuration  are  shown  in  the  following
figures.  The  communication  port  mating  connector  is
supplied  with  the  RS  232  PCB  (AMP/Cannon  or
equivalent 9-pin "D" type male connector). The external
cable (not supplied) between the host computer and the
controller does not require crossed wires so that signals
are connected correctly .
For  example,  the  Transmit  data  signal  from  controller
(pin  2)  must  be  connected  to  the  host  computer's
Receive  data  line  (pin  2)  and  vice  versa.  Consult  the
host  computer's  instruction  manual  for  its  serial  port
connections

NOTE

Varian  cannot  guarantee  compliance  with  FCC
regulations  for  radiated  emissions  unless  all  external
wiring  is  shielded,  with  the  shield  being  terminated  to
the  metal  shroud  on  the  O-subconnector.  The  cable
should be secured to the connector with screws.

 

!

In order to avoid possible conflicts on the Serial Line, it
is advisable to use a 3 wire shielded cable for the TxD,
RxD  and  GND  connections  and  to  leave  all  the  other
pins unconnected.

RS 232 Communication Descriptions

Communication RS 232 serial port connections

Transmission Channel Characteristics

levels: 

RS 232/RS 422

baud rate: 

9600/4800
programmable by a jumper on the 
board

character length:  8 bits
parity: 

none

stop bit: 

1 bit

protocoll: 

master (PC) / slave (converter)

In this case the value to be assigned to the ADDRESS
field must be 80 hex (for RS 232).

Message Structure

(request and answer have the same format)

The  master  system  (PC)  starts  every  session  sending
the following message to the slave units connected:

<STX> / <ADDR>  + <WINDOW> + <COMMAND> +
<DATA> + <ETX> + <CRC>

where:

<STX>=

0x02

<ADDR> =

0x80 (for RS 232 and
RS 422 only)

<ADDR> =

0x80 + device number (0...31)

0xFF: brodcasting command 
(recognized by all the devices, it 
doesn’t implicate any answer)
(for RS 485 only)

<WINDOWS>=

'000'...' 999' window number
the meaning of the window
depends to the device type

<COMMAND>=

0x30 :window value reading
0x31 :window writing

<DATA> =

alphanumeric ASCII string
containing, in the case of writing
operation, the parameter to input
into  the window  addressed by
the field <WINDOW>This field
may have variable length
according to the data type
contained in the window where
you are working in. In the case of
reading request of a window, the
data field doesn’t exist.

<ETX>=

0x03

<CRC>=

XOR among all the characters
following <STX>=(with exception of
<STX>), including the end
character <ETX> hexadecimally
encoded by two ASCII characters.

 

WARNING!

Summary of Contents for Turbo-V 300 75 Vdc

Page 1: ...87 900 939 01 D APRIL 2002 INSTRUCTION MANUAL Turbo V 300 75 Vdc Box Controller Model SQ 189 vacuum technologies...

Page 2: ...Turbo V 300 75 Vdc Box Controller...

Page 3: ......

Page 4: ......

Page 5: ...ump Connector 4 I O Specifications 4 9 pin D Type Connector Pin Assignement 4 Serial Communication Port 5 RS 232 Communication Descriptions 5 Transmission Channel Characteristics 5 Message Structure 5...

Page 6: ...stand operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maint...

Page 7: ...and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output voltage according to the software and the gas load condition of th...

Page 8: ...TECHNICAL INFORMATION 3 87 900 939 01 D CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure Controller outline...

Page 9: ...ins A F Pump temperature sensor Pin E Ground I O Specifications START STOP START command STOP command Low 0 8 Vdc High 4 to 15 Vdc Analog output Output impedence Minimum load 0 to 10 Vdc proportional...

Page 10: ...ommunication RS 232 serial port connections Transmission Channel Characteristics levels RS 232 RS 422 baud rate 9600 4800 programmable by a jumper on the board character length 8 bits parity none stop...

Page 11: ...35 STX ADDR ACK ETX CRC Command SOFT START ON Source PC Destination Inverter 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Inverter Destination PC 02 80 06 03 38 35 STX ADDR ACK E...

Page 12: ...4 brake 5 normal 6 failure 206 N X ERROR CODE Too high load No connection Short circuit Pump overtemp SoftStart Not Ended Controller overtemp RunUpTime Not Reached Power fail 211 N X PUMP SENSOR TEMPE...

Page 13: ...9 pin D type connector To deactivate the Low Speed status it is necessary to disconnect pin 9 from pin 15 ground of the same connector The low speed frequency is equal to 622 Hz MAINTENANCE Replacemen...

Page 14: ...TECHNICAL INFORMATION 9 87 900 939 01 D PCB JUMPERS...

Page 15: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Page 16: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Page 17: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...