ProStar 325 UV-Vis Detector Control Manual
-
29 -
03-914947-44:2
The Peaksense relay is only activated in the following states:
ready, running, and stopped. When the acquisition reaches the
end time Time Slice function is turned off.
To assign the correct value to the Period parameter measure
the peak width at half height of the peak or peaks of interest.
Divide this value by the number of segments you want to
collect, and enter that value.
Pulse Relay Function
When this function is chosen, a single peak sensing relay
activation takes place at the specified times. It is used to send a
single relay closure to the sample collector at a specified time
for any purpose you may determine, (e.g., a user would
estimate that the sample receiver will fill and overflow at about
three minutes, the user may set the Pulse Relay function to
advance the sample collector at 2.75 minutes to the next
sample receiver).