CP-4900 Micro-GC
Pre-installation requirements
Varian, Inc.
Part number: CP501289
Page: 7
Operation after Long Storage
Follow the procedure below if your Micro-GC has been stored for a long period!
•
Put carrier gas on the Micro-GC
•
Switch the Micro-GC ON
•
Switch the TCD filament(s) OFF
•
Set the column(s) temperature to it’s maximum (depending on the column module)
•
Condition the column module, preferably overnight. This will ensure you that all the
water has been removed from the column module and no damage will occur
to the TCD filaments
Outputs
The Micro-GC has different output ports for interface with a computer system.
Digital: 1x RS-232 port, 1x Ethernet port, 2x RS-232 ports for external events.
Network Requirements of the CP-4900 Micro-GC
•
Customer Cable type should be Cat5 UTP / STP.
•
Customer network should comply with Standard Ethernet (IEEE 802.3).
•
The GC can be plugged into any 10BASE-T or 10/100BASE-TX compatible HUB or
Switch.
•
TCP-IP should be used on the Network.
Minimum Computer requirements for all CP-4900 Micro-GC systems
•
Pentium processor > 1.5 GHz.
•
Ram > 512 MB
•
Hard disk space > 20GB
•
Windows
XP
•
1 USB port
•
1 Serial port