4-2 Error messages
For a certain type of failure, the controller will self-
diagnose the error and the following messages will be
displayed.
The controller signals the error occurred by means of a
diagnostic LED located on the board, and on the
RS 232 port.
The LED blinks in a coded mode:
it flashes a number of time equal to the error
code (see the following table) and then stays off, and
so on.
Error Code Table
LED
BLINKING
NUMBER
DESCRIPTION
0
No error
1
Output current> 15 A
2
Not connected pump
3
Output current > 1.5
A for 15 sec.
4
Bearing temperature > 60 °C
5
Heat exchanger temperature
> 60 °C
6
Normal not reached at run up time
7
Input voltage < 16 V for 5 sec.
8
Operating voltage < 10 V for 5 sec.
9
Vp voltage 8V lower than nominal for 5
sec.
10
Soft start ramp not ended within the
expected time