Models 947, 948 and 960 Leak Detectors Operations and Service Manual
B-2
B.1.1 Opening the Turret
The turret opens by unlocking and lifting its cover. To remove the top cover, you must
remove the shelf that the service pendant sits on.
B.1.1.1 Removing the Service Pendant Shelf
To remove the service pendant shelf:
1. Loosen the screws of the service pendant serial connection and remove the connection
NOTE
The service pendant power cable is still attached to the
electronics housed beneath the pendant shelf.
2. Place the service pendant on the top of the 947/948/960 or on another nearby surface.
Figure B-1
Removing the Service Pendant Shelf
3. Use a Phillips head screwdriver to remove the screws located at either side of the shelf.
4. Lift the shelf up and out of the turret.
Serial
Screw
Screw
Connection