Saturated filters on the GC may produce an increase in the air/water background.
Replace the filters at regular intervals, and whenever moisture or other
background from the GC becomes a problem.
GC Troubleshooting
NOTE: Please refer to the GC Operator’s Manual for information about GC
troubleshooting and diagnostics procedures not described in this section.
This section describes chromatographic troubleshooting, with particular
emphasis on GC/MS applications. You will be able to investigate most of the
problems addressed in this section by running the COLTEST mixture
(03-920273-00).
The following procedure describes the chromatographic conditions and the
expected results when running the COLTEST sample with a 30-m vf5ms column
(0.25 mm ID, 0.25 µm film thickness).
Using the COLTEST Sample for Troubleshooting
The Coltest sample provides a good mechanism for identifying a variety of
chromatography problems. A COLTEST method can be found in the
C:\VarianWS\4000 MS Service directory of the software.
Running the COLTEST Sample
Flow Pressure Conditions
Use a constant flow of 1.0 mL/min.
Injector Conditions
•
1177
Injector:
Use an isothermal temperature of 240 °C.
Set up the following split program conditions:
Time
Split State 1
Split Ratio
initial On
100
0.01 Off
Off
1.00 On
100
MS Temperature Conditions
1. Set the transfer line temperature to 250 °C.
2. Set the trap temperature to 150 °C.
3. Set the manifold temperature to 40 °C.
4. Set the source temperature to 150 °C if the 4000 MS is in external mode.
109 of 127
Summary of Contents for 4000 GC
Page 4: ......
Page 40: ...6 of 127...
Page 77: ...3 Check source connection pins for proper alignment and straighten as necessary 43 of 127...
Page 124: ...90 of 127...
Page 148: ...114 of 127...