5. Power and Environmental Specifications
Immunity test
EN/IEC 60601-1-2
test level
Compliance level
Electromagnetic
environment — guidance
Voltage dips/dropout
EN/IEC 61000-4-11
0 % Dip; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0 % Dip; 1 cycle
70 % Dip; 25 cycles
for 50Hz at 0°
70 % Dip; 30 cycles
for 60Hz at 0°
0 % Interruption;
250 cycles for 50Hz
at 0°
0 % Interruption;
300 cycles for 60Hz
at 0°
0 % Dip; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0 % Dip; 1 cycle
70 % Dip; 25 cycles
for 50Hz at 0°
70 % Dip; 30 cycles
for 60Hz at 0°
0 % Interruption;
250 cycles for 50Hz
at 0°
0 % Interruption;
300 cycles for 60Hz
at 0°
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the system requires continued
operation during power mains
interruptions, it is recommended
that the system be powered from
an uninterruptible power supply or
battery.
Power frequency 50/60
Hz magnetic field EN/
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
Proximity Magnetic
Fields IEC 61000-4-39
134.2 kHz at 65 A/
m, 13.56 MHz at 7.5
A/m
134.2 kHz at 65 A/
m, 13.56 MHz at 7.5
A/m
RFID equipment should be
used no closer to any part of the
device, including cables, than the
recommended 15 cm separation
distance
Table 9: Emissions — All equipment and systems not life-supporting
Emissions test
EN/IEC
60601-1-2
test level
Compliance
level
Electromagnetic environment — guidance
Conducted
RF EN/IEC
61000-4-6
3 V 0.15 MHz –
80 MHz,
6 V in ISM and
Amateur bands
between 0.15
MHz and 80
MHz, 80 % AM
at 1 kHz
(V1)=3 V 0.15
MHz – 80 MHz,
6 V in ISM and
Amateur bands
between 0.15
MHz and 80
MHz, 80 % AM
at 1 kHz
D=(3.5/V1)(√P)D=(3.5/E1)(√P) 80 to 800 MHz
Radiated RF EN/IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
(E1)=3V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
D=(7/E1)(√P)
800 MHz to 2.7 GHz
Where P = max power in watts and D =
recommended separation distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance levels (V1 and
E1).
Interference may occur in the vicinity of equipment
containing a transmitter.
Nexus DR Regulatory and Safety Manual, 736-226-G1, Rev. J | 27
Summary of Contents for nexus DR
Page 1: ...Regulatory and Safety Manual 736 226 G1 Rev J...
Page 2: ...Regulatory and Safety Manual 2 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 13: ...1 Regulatory Information 4 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 17: ...2 Statement of Intended Use 8 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 21: ...3 Cleaning Practices 12 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 44: ...Index 35...
Page 45: ......