Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09
17/24
www.vareximaging.com
8.
Disposal, ESD and EMC compatibility
Disposal
This device contains substances that can be hazardous to the environment and care should be taken
when disposed of.
The device is marked with the following symbol:
Follow local regulations regarding disposal of devices that contain electronic parts.
ESD
WARNING:
The device contains sensitive electronics. Ensure that ESD protective measures are in
place when the device is installed or serviced to prevent damage to the device.
EMC compatibility
The device conforms to IEC 60601-1-2:2014 for EMC compatibility and must be installed and put into
service according to the EMC information provided in this manual.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the AmpMC 3-field, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
WARNING:
Not taking EMC measures into account on the wiring may result in increased
EMISSIONS or decreased IMMUNITY. IEC60601-1-2:2014 must be followed for being
complaint with EMC guidelines.
WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased Electromagnetic
emissions or decreased Electromagnetic immunity of this equipment and result in
improper operation.
NOTE:
The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Deviations
No deviations from IEC60601-1-2:2014 are applied.