Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09
8/24
www.vareximaging.com
Restrictions on use
The AmpMC can only be used in combination with a SolidStateMC and in a Radiography X-ray system
that complies with the IEC60601-1 standard, applicable at date of manufacture.
The supply lines must be limited / fused to 15W max. or the AmpMC must be mounted inside a fire
enclosure.
The AmpMC is intended to be installed inside a cabinet of the X-ray system.
Depending on the final assembly, additional measures may need to be taken to comply with EMC
regulations.
Contraindications
The relevant contraindications for the X-ray system continue to apply (see the documentation for the
X-ray system). The AmpMC does not add new contraindications on top of these.
Overview of the device
Table 1 AmpMC specifications
Model
Dimensions
(mm)
Support information
Lay
-out
Remarks
Outside
#
of
fi
el
ds
Connections
SolidStateMC
Generator
W
L
H
Cable
(m)
Connector
Cable
(m)
Connector
1001
42 222 21
3
None 6 pin Female 0.93 Sub-D 9p male
§ 6.1 3-field SolidStateMC preamp
with single voltage ramp output
1006
3-field SolidStateMC preamp
with single voltage ramp output
and modified reset circuitry
1008
0.8 AMP Header 10p female
/ GND terminal
3-field SolidStateMC preamp
with single voltage ramp output
Configuration of connections
Connector type at generator side
Cable length at generator side
Connector type at SolidStateMC side
Cable length at SolidStateMC side
# of fields that can be controlled
Height of AmpMC
Length of AmpMC
Width of AmpMC