2315NDT v2 CMOS X-ray Detector User Manual
11.0
Standards, Directives and Regulations
The 2315NDT X-ray detectors are components designed to be compliant with the standards,
directives and Regulations detailed in
,
to support system compliance when incorporated
into X-ray equipment.
12.0
Storage and Use
2315NDT flat panel X-ray detectors can be used or stored within the conditions listed in
Table 5
Standards, Directives and Regulations
Standard/Directive
Description
EN/IEC61326-1:2012
Electrical Equipment for Measurement, Control and Laboratory
Use - EMC Requirements - Part 1: General Requirements
EN/IEC61010-1:2010
Safety requirements for electrical equipment for measurement, control
and laboratory use – Part 1: General requirements
2011-65-EU
a
+2015/863
a. Detectors employ Lead as exempted under Annex III Item 13(a) and Annex IV Item 5.
Restriction of hazardous substances (RoHS)
EC-1907/2006
b
b. Per Article 33, Detector components contain Lead Oxide.
Registration, evaluation, authorization and restriction of chemicals
(REACh)
Table 6
Storage and Use Conditions
Operation and Storage (without Packing)
Operating temperature
+10°C to +40°C
Storage temperature
+10°C to +40°C
Humidity
10% to 80% RH, non-condensing
Atmospheric pressure
690 mBar – 1080 mBar
Transportation (Packaged)
Temperature range
-10°C to +45°C for 30 days
Humidity
10% to 90% RH, non-condensing
Atmospheric pressure
690 mBar – 1080 mBar
WARNING
When Varex Imaging flat panel X-ray detectors are transferred from
extreme temperature or humidity conditions, allow at least one hour of
stabilization to normal operating temperature and humidity before
use.