DISPOSABLE PATIENT CIRCUIT

START UP AND ADJUSTING PARAMETERS

CONNECT TO PATIENT

INTERNAL BATTERY BACK-UP

SHUT DOWN

May be patented. www.vapotherm.com/patents

Technical Support

T: 855 557 8276 Domestic 

+1 (603) 658-5121 International

[email protected]

•  Press the Run/Standby button. Unit will enter Standby mode (No Flow), indicated by 

the AMBER light

•  Clamp the water inlet tube. Open the door, remove the Disposable Patient Circuit 

(includes delivery tube & cartridge) by sliding it upwards out of the docking station

•  Discard all disposables according to hospital guidelines

•  Disconnect unit from AC power

•  Wipe down with Super Sani-Cloth®. In addition, if hospital procedures require, the 

following may be used: 70-90% Isopropyl Alcohol, 2% (maximum) chlorine cleaning 

solution, 6% (maximum) Hydrogen Peroxide cleaning solution, CaviWipes

TM

, AF3 

Germicidal, Incidin® OxyWipe, Bacillol® 30 Tissues, Clinell® Alcohol Wipes, or Tuffie 

Disinfectant Wipes

•  The back-up battery is designed for temporary use only, when AC power to the unit 

has been interrupted

•  The unit will enter into battery mode and will maintain flow and oxygen percentage 

for at least 15 minutes

•  The battery icon will flash

•  Replace battery every two years

•  Battery recharges in two hours

•  The flashing green LED becomes steady when the set variables are reached

•  Place the cannula on the patient. Once the unit has reached at least 33°C, connect to 

delivery tube

•  The unit should not be placed in Standby mode for extended periods of time. For 

pauses in therapy, keep unit in RUN mode, remove cannula from the patient, and set 

the parameters to the lowest available setting. To reinitiate therapy, before cannula is 

placed on patient, clear accumulated condensate

•  Install oxygen sensor. Replace sensor annually

•  Install gas inlet filters on back with filter bowls vertical (glass side down). Replace gas 

inlet filters every six months

•  Attach heliox & O₂ hoses. Plug in power cord

•  Rotate the blue Setting Control Knob to illuminate display

•  Press in Setting Control Knob to select the parameter and rotate to adjust the value 

•  Press and release (do not hold) the Run/Standby button once to start 

A GREEN light indicates RUN mode (AMBER light indicates STANDBY mode, No Flow) 

•  Green light will stop flashing once temperature is reached

•  Assemble Disposable Patient Circuit according to provided instructions.

•  Open door and install Disposable Water Path into docking station so there is no gap 

between bottom of the Disposable Water Path and the Docking Station Floor

•  Hang sterile water bag or bottle

•  Wipe water spike with alcohol pad or equivalent and insert into sterile water bag

•  Allow a minimum of 200ml of water to fill into the Disposable Patient Circuit

•  Precision Flow Heliox is ready for start up

0297

RMS-UK Ltd.

28 Trinity Road

Nailsea, North Somerset

BS48 4 NU

United Kingdom

Summary of Contents for Heliox Precision Flow

Page 1: ...tridge fault Gas supply fault Run Stop Status LED Run Standby button Q U I C K R E F E R E N C E G U I D E High Flow Circuit Installed This guide provides you with basic instructions on how to set up and operate the Precision Flow Heliox Before operating the Vapotherm Precision Flow Heliox please review the Instructions For Use which can be found at our website www vapotherm com and on the USB dri...

Page 2: ...ADY The internal battery backup is not fully charged The unit would not run on battery for the full rated time in the event of a power failure No action is necessary BATTERY FLASHING The unit is running in BATTERY mode Gas flow and blending continues without heat or water circulation AC power is disconnected Reconnect AC power BLOCKED TUBE FLASHING High back pressure Obstructed or kinked cannula d...

Page 3: ...previous temperature Silence alarm and wait for temperature to drop Temperature 2 set point Very low water temperature after bag replacement Silence alarm and wait for temperature to rise INDICATIONS WARNINGS AND CAUTIONS Primary Indications The Precision Flow Heliox is intended to add warm moisture to breathing gases from an external source for administration to a neonate infant pediatric and adu...

Page 4: ... when the set variables are reached Place the cannula on the patient Once the unit has reached at least 33 C connect to delivery tube The unit should not be placed in Standby mode for extended periods of time For pauses in therapy keep unit in RUN mode remove cannula from the patient and set the parameters to the lowest available setting To reinitiate therapy before cannula is placed on patient cl...

Reviews: