2.

 

System Setup 

2.1

 

Installation 

2.1.1

 

Remove the RF Generator and all other Vante™ components from their 

respective shipping cartons and visually inspect them for obvious damage.  
Contact Vante™ if any damage is found. 

2.1.2

 

Check to see that the power cord plug matches the power receptacle for the 

country in which the system is being used.  If it does not, contact Vante™.  

CAUTION:

  Do not connect the power cord to AC mains at this time.

 

2.1.3

 

Follow the separate installation instructions provided with the unit for this 

application. 

2.1.4

 

Verify that all connections are correct before proceeding. 

2.2

 

Adjustment and Operation 

WARNING:

 

Do NOT place fingers or any other objects near or between the 

electrodes in the forming/welding head when the generator RF 
power is on.  Severe burns or damage to the electrodes can 
occur!

 

2.2.1

 

Verify the AC power switch is in the “off” position and connect the AC 

power cord to any nominal ac voltage in the 100 - 230 Vac range. 

2.2.2

 

Turn the RF Generator on by toggling the rocker switch located on the back 

of the unit.  The power-on light will illuminate green, indicating that power is on.   
The power-on light will be amber when RF energy is being produced.  The fan in 
the back of the Generator will begin to operate.  The Generator requires no warm-
up period due to transistorized electronics design. 

NOTE:

 

Vante™ may recommend initial control settings for start-up purposes 
only; however, the user is responsible for determining the proper 
control settings for particular applications. 

2.2.3

 

Set the control settings and air pressure to the appropriate level for the 

desired application.  RF dwell time should be no longer than necessary to make a 
good seal or weld. 

Summary of Contents for 3120

Page 1: ...Doc No 31200801 00 Rev A INSTRUCTION MANUAL MODEL 3120 200 Watt RF Generator...

Page 2: ...ii This page intentionally left blank...

Page 3: ...5 10 Vante iii Vante Engineering Research Associates Inc 100 N Tucson Blvd Tucson AZ 85716 USA Telephone 520 881 6555 Fax 520 323 9055 Toll Free 877 565 5557 Part No 31200801 00 Rev A...

Page 4: ...iv This page intentionally left blank...

Page 5: ...ed by a sealing head such as the Model 3806 mid size Tube Sealer for making seals on tubing made of RF reactive thermoplastic materials such as polyurethane polyvinylchloride PVC Ethylene vinyl acetat...

Page 6: ...e assumes no liability or obligation nor guarantees product performance Proprietary Information All rights are reserved Copying of the protected designs associated with the Model 3120 is strictly proh...

Page 7: ...ystem 1 1 2 Theory of Operation 1 1 3 User Safety 2 1 4 Controller Interface 2 1 5 Specifications 3 1 6 Component Identification 4 2 System Setup 7 2 1 Installation 7 2 2 Adjustment and Operation 7 3...

Page 8: ...viii Table of Figures Table 1 1 Control Port Pin Out 3 Figure 1 1 Model 3120 RF Generator Front View 4 Figure 1 2 Model 3120 RF Generator Rear View 4 Figure 1 3 Serial Plate 5 Figure 1 4 Fuse Labels 5...

Page 9: ...may be integrated into the forming or welding head 1 1 4 The head itself may be designed and manufactured by Vante or the customer may design and manufacture his own in close association with Vante s...

Page 10: ...2 Vante RF Sealers meet or exceed appropriate electrical safety standards and pose no electrical shock hazard when used with properly fused and grounded outlets Reference Radio Frequency System Safet...

Page 11: ...5 Specifications 1 5 1 Mechanical Size H x L x W 4 in x 17 in x 12 in 10 cm x 43 cm x 30 5 cm Operating Weight 27 lbs 12 3 kg Shipping Weight 30 5 lbs 13 8 kg 1 5 2 Electrical Power Input 100 230 VAC...

Page 12: ...s for the model Each unit s serial plate indicates the model number and serial number of the unit 1 6 2 Refer to the serial number when contacting Vante or one of its authorized service centers BNC Co...

Page 13: ...Vante Model 3120 5 Figure 1 3 Serial Plate Figure 1 4 Fuse Label...

Page 14: ...Vante Model 3120 6 This page intentionally left blank...

Page 15: ...en the generator RF power is on Severe burns or damage to the electrodes can occur 2 2 1 Verify the AC power switch is in the off position and connect the AC power cord to any nominal ac voltage in th...

Page 16: ...e recommended setting should be required 2 2 6 Once the desired result is obtained continue to perform the sealing forming or welding operation as needed NOTE The RF Generator is equipped with a therm...

Page 17: ...luminated when the rocker switch is in the on position then one of the fuses may need replacement 3 2 2 To access the fuse compartment the AC cord must be unplugged Release the fuse holder slide latch...

Page 18: ...Vante Model 3120 10 This page intentionally left blank...

Page 19: ...rgy However misuse or direct contact between tissue and RF electrode s can result in severe RF burns 4 3 RF Effects On Pacemakers 4 3 1 There is no evidence that Vante RF instruments interfere with th...

Page 20: ...n any area with a potentially explosive atmosphere It is possible for the RF electrodes to arc initiating an explosion or fire NOTE Vante recommends strict adherence to the procedures specified in the...

Reviews: