background image

 

© October 2021 Van Essen Instruments. All rights reserved. 

www.vanessen.com 

Diver-Link Preparation 

Step 5   Configure the Diver-Link Hardware 

Insert the SIM card, 1 or 2 batteries and desiccant in the Diver-Link and connect the external antenna 
(if applicable). 

 

Carefully note the polarity of the batteries: these must be 
inserted in parallel, facing the same direction. Polarity is 
indicated on the bottom of the battery holders.

 

 

Ensure no wires are pinched between the lid and the base 
when closing the box. 

 

A Diver-Link with an external antenna (DN402, DN422, DN402-3, DN422-3) will not transmit 
data if the external antenna is not connected to the Diver-Link. 
 

 

Diver-HUB Getting Started Guide

, sections 2.2.1 Opening the Diver-Link, 2.2.2 Inserting a SIM 

Card, 2.2.3 Inserting the Battery, 2.2.4 Inserting the Desiccant and 2.2.5 Closing the Diver-Link. 

Step 6   Configure the Diver-Link in Diver-HUB 

Configure the Diver-Link in Diver-HUB. Go to Equipment Management Diver-Link Configuration. 

 

This will configure all Diver-Links (transmission interval, time of transmission) for a given 
project in Diver-HUB. 
 

 

Diver-HUB Getting Started Guide

, section 4.2.2 Diver-Link Configuration. 

Setup Divers 

Step 7   Import Diver-HUB Project in Diver-Office  

Import Diver-HUB project in Diver-Office program by using your Diver-HUB credentials.  
Go to Project > Diver-HUB > Import Project Data 
 

 

Once the Diver-HUB project has been imported into a Diver-Office project, the same Diver-
Office project must always be used to program any Diver for that project, also after 
installation. For example, when reprogramming or replacing a Diver. A new import from 
Diver-HUB is required when changes have been made to the Diver-HUB project. For example, 
when new locations/monitoring points were added. 
 

 

Diver-HUB Getting Started Guide

, section 4.2.3 Diver Configuration. 

Step 8   Program the Divers 

Program the Divers using Diver-Office. A Diver can be programmed with a USB Reading Unit or with a 
Smart Interface Cable connected to a DXT Cable and Diver 

 

Select the Monitoring Point name from the Monitoring Point dropdown list to avoid mistakes. 

 

Make sure to also start the Diver or program a future start date. 

 

 

Diver-HUB Getting Started Guide

section 4.2.3 Diver Configuration.

 