background image

V&P Scientific, Inc.  

Technical Note 104A 

2)

 

Turn the motor drive shafts so the two coupler prongs align with the grooves in the Bubble Paddle 
couplers on the VP 759-1 Reservoir (see Figure 4, page 2). Line up the reservoir with the base and 
slip the Bubble Paddle couplers onto the motor drive shaft couplers (see Figure 5a, 5b and 5c below). 
It helps to lift up slightly on the opposite end of the reservoir while pushing forward (see Figure 5c 
below). Once the couplers are engaged lower the reservoir so that the two alignment pins in the 
base are in the holes in the underside of the reservoir (see Figure 3, page 2). 

 

 

 

 

Figure 5(a-c).  Engaging couplers of VP 759-1 Direct Drive Bubble Paddle Reservoir  

and VP 769-1 Motor Base. 

 

3)

 

Connect all of the cables as shown in Figure 6 (page 4).  Make sure that On/Off switch is off when 
connecting  power  supply.  Never  directly  connect  power  supply  to  motor  base.  Always  connect 
through the control box. 

5c 

5b 

5a 

Summary of Contents for VP 769-1 Series

Page 1: ...ive Bubble Paddle Reservoir Figure 1 VP 769 1 Motor Control Unit shown with VP 759 1 Direct Drive Bubble Paddle Reservoir Figure 2 Close up view of the Direct Drive Bubble Paddle Reservoir VP 769 1 Mo...

Page 2: ...VP 769 1 drive shaft couplers Since the splash guard is raised it is best if there is no liquid in the reservoir Figure 4 Couplers of VP 759 1 Direct Drive Bubble Paddle Reservoir and VP 769 1 Motor B...

Page 3: ...n the opposite end of the reservoir while pushing forward see Figure 5c below Once the couplers are engaged lower the reservoir so that the two alignment pins in the base are in the holes in the under...

Page 4: ...osition the pipet tips in alignment with the notches in the Bubble Paddles and the depressions in the bottom of the Bubble Paddle Reservoir see Figure 2 page 1 6 To assist in this alignment with a dec...

Page 5: ...type of solution and volume being stirred A short warm up period 30 min is recommended For most applications a speed setting of 20 on the control dial is optimum 400 RPM See chart below of Speed in RM...

Page 6: ...or Isopropanol for easy drying 3 Removing and replacing Bubble Paddles can be done for cleaning or if paddle has been damaged a Remove VP 759 1 series Reservoir from the VP 769 1 Motor Base as descri...

Page 7: ...ve shaft and move slightly up and down until hub end falls into place g Push the drive shaft back into the sealed bearing and turn the locking screw to secure the location of the sealed bearing to the...

Page 8: ...breakdown may result WARRANTY V P Scientific Inc warrants this product to be free from defects in material and workmanship when used under normal laboratory conditions for one year This warranty begi...

Reviews: