background image

V&P Scientific, Inc. 

Technical Note 380A 

4. Connect power cord into the power receptacle on the back of the Temperature Controller.  Connect the 

power cord to a 120 V outlet (VP 741D).  In the case of CE compliant models, connect the power cord to a 
230 V outlet (VP 741DCE). 

 

OPERATION 

HEATING BLOCK OPERATION USING THE TEMPERATURE

 

CONTROLLER 

1. Turn on the Temperature Controller using the on/off power switch.  The Controller already has a temperature 

programmed so the Heating Block will immediately begin to heat.  The red number in the upper display is the 
current temperature of the Heating Block.  The green number in the lower display is the set point during 
operation.  See the Temperature Controller Operating Instructions (Appendix) for more details on features of 
the display. 

 
2. To turn off the Heating Block without turning the power switch off, press the down arrow key until green lower 

display reads “OFF”.

 

 

 

Figure 3. Temperature Controller Display for the VP 741D Series Heat Block 

 

3. Although the center of the Heating Block will be at the temperature on the display, the actual temperature of 
samples in the microplate wells or tubes will be 3°C to 5°C below this temperature depending upon the efficiency 
of  heat transfer.   It is  recommended  that the temperature  in  one  or  more  samples  be  measured  directly  to 
determine appropriate setting for a particular application.  

4. For faster heat transfer, use a Heating Block Insert such as one that conforms to the bottom of a plate (for 
example, VP 741I6B or VP 741I6C) or an insert that holds tubes or vials.  Contact V&P Scientific for information 
on these items. 

5. The upper limit temperature on the Temperature Controller is set to 120

o

C at the factory.  Temperatures higher 

than  120

o

C  should  not  be  used  with  polypropylene  tubes  or  microwell  plates.    Polypropylene  will  melt  at 

temperatures  greater  than  130

o

C.    To  operate  at  temperatures  above  120

o

C,  use  of  glass  or  other  high 

temperature-stable containers is recommended. 
 

 

 

Upper 

Display 

Lower 

Display 

Up/Down 

Arrows 

On/Off 

Power Switch 

Set 

Button 

Shift Button 