Installation and Operation Instructions
GoLift Service and Repair Manual

Summary of Contents for GoLift

Page 1: ...Installation and Operation Instructions GoLift Service and Repair Manual ...

Page 2: ... GoLift Engaging The Trolley Micro switch Replacing The GoLift Strap Replacing The GoLift Battery PCB Connections Replacing the PCB Replacing the GoLift Hand Control Replacing The Emergency Switch Replacing the GoLift Motor s Checking The Overspeed Cam Replacing the Limit Switch Spare Parts Warranty 19 ...

Page 3: ... manual patient transfer and repositioning which has led to the enforcement of more strict lifting regulations Vancare strives to consistently provide exemplary service design and installation to all of its customers In an ever more competitive market it is important to standardize installation techniques thus ensuring that the highest standards of quality and safety are met The methods described ...

Page 4: ... upside down to the side showing the battery compartment lids 3 Twist the knobs to unlocked position and remove the battery lids 4 Use a 5 32 T Handle to remove the two screws fastening the cover 5 Position the battery lids 6 Turn the knobs to locked position Getting Access to the GoLift ...

Page 5: ...Getting Access to the GoLift 7 Turn the lift upside down and gently grasp the bottom cover and pull upwards 8 Set the bottom cover aside In order to operate the lift with the hand control the trolley needs to be inserted into the lift Another method to allow the lift to operate is plugging in a jumper connector in the limit switch port in the circuit board ...

Page 6: ...ase the complete length of strap Replacing The GoLift Strap Engaging The Trolley Microswitch Alternatively a jumper can be used to activate the trolley micro switches in the GoLift 2 Disconnect the trolley micro switch harness and plug in the jumper in the proper circuit board port ...

Page 7: ... hub assembly 5 Pull the strap out of the gearbox assembly 6 Insert the new strap with the correct orientation as shown through the opening in the bottom cover of the GoLift 7 Position the strap through the limit switch rollers 8 Place the strap in the opening of the strap hub 9 Insert the shoulder bolt and tighten using the 5 32 T Handle Replacing The GoLift Strap ...

Page 8: ...ll the emergency red power cord to ensure all power is cut off from the battery 3 Open the battery lids to expose the batteries 4 Grasp the connector and pull outwards to disconnect the batteries 5 Install new batteries in the compartments and plug in to the respective PCB ports ...

Page 9: ...www vancare com 9 PCB Connections Battery Connectors Charger Emergency Switch Hand Control Motors Latches Motor Brake Limit Switch ...

Page 10: ...D equipment wrist strap prior to handling the PCB 2 Open up the GoLift by following the steps in the section Getting access to the GoLift 3 Unplug all the harnesses from the PCB 4 Use a 3mm T Handle to remove the two 2 screws fastening the PCB to the gearbox ...

Page 11: ...eplacing the PCB 5 Position the new PCB on to the top cover of the GoLift Make sure to line up the connector with the opening in the cover 6 Plug in all the harnesses 7 Fasten the PCB to the top cover using a 3mm T handle ...

Page 12: ...ion Getting access to the GoLift 2 Disconnect the hand control harness 3 Twist the hand control grommet until it is able to pass through the opening in the cover 4 Fish the remaining cord and connector through the opening in the cover 5 Install the new hand control by following these steps in reverse ...

Page 13: ...re com 13 Replacing the Emergency Switch 1 Use a Phillips screwdriver to remove the two 2 screws securing the emergency stop switch bracket 2 Once the bracket is removed the emergency stop sleeve can be removed ...

Page 14: ...14 Vancare Replacing the Emergency Switch 3 Use an 8mm socket wrench to remove the emergency switch from the bracket 4 Disconnect the emergency switch 5 Use cable tie cutters to remove the cable ties ...

Page 15: ...e GoLift Motor s 1 Use a 3mm T Handle to remove the three 3 screws securing he motor brake 2 Disconnect the motor harness from the motor 3 Use a 3mm T Handle to remove the three 3 screws securing the motor to the motor mounting plate ...

Page 16: ...16 Vancare Replacing the GoLift Motor s 4 Slide the motor out of the motor mounting plate 5 Install new motor and motor brake following the previous steps in reverse ...

Page 17: ... 2 Use a plastic mallet and a bar to push the gear hub and gear out 3 Use a T handle or a small screw driver to confirm that the overspeed cam is operating correctly and not getting stuck 4 Reassemble the gear hub and gear by pushing the gear back into place towards the plate and reinserting the retaining ring ...

Page 18: ...to remove the limit switch harness in case of limit switch failure 2 Install a new limit switch harness and test it to confirm that these are operating correctly LIFT QR TRLY ASSY EL BAT PACK AA 2AH LIFT MTR ZN1203 EL CRC BRD MMB GL LIFT STRAP GO LIFT HND CTRL 2 Spare Parts ...

Page 19: ... installation costs after the first twelve 12 months will be borne by the customer The following exclusions apply the warranty for batteries is for a period of three 3 months from the time of installation the warranty for power supply is one 1 year from the time of installation This warranty is valid only when the product has been properly installed as outlined in the Vancare specifications includ...

Page 20: ...Vancare www vancare com 1515 1st Street Aurora NE 68818 Toll Free Tel 1 800 694 4525 Tel 402 694 4525 Fax 402 694 3994 Email Info vancare com www vancare com ...