49
4.4 Probe maintenance
►
Probe is an expensive part and frangible. Never hit it or drop it on floor.
►
Please use medical ultrasound gel during diagnoses. Do daily inspection on the probe enclosure to see if it is cracked
and avoid liquid leakage to spoil the built-in components
►
Check the probe regularly to see if it is filled with clean medical castor oil lest the existing air bubbles influence the
image quality. If air bubbles appear, turn the probe oil filler hole up and screw off the sealing screw, turn the probe
slowly to drive the air bubbles to the screw hole, infuse a little bit of castor oil into the screw hole with syringe with a
needle (refer to Picture 6-1), vent the air bubbles and fasten the sealing screw and finally clean the oil stain on the
probe surface.
Oil injection hole Probe chip
Scanning direction
Mechanical sector scan probe oil injection, and scanning direction
4.5 Cleansing
When the enclosure need cleaning, wipe it with soft dry cloths and then wipe gently with sponge dipped
with 75% medical alcohol.
Warning
:
To prevent accidents, take out the battery when cleaning the main unit enclosure and separate the device from
the power supply network first and then clean the adapter enclose.
Prevent all the plugs, sockets from water splash or socking.
Warning
: