VALTRONICS 2011 Instruction Manual Download Page 9

 
 

Copyright 2011 VTI-Valtronics Inc. 

 

 

 

                                           
                                          Calibration 
 
 
 

   
                                                               

CO

2  

Calibration

 

 
The meter may be calibrated with either fresh air or with the optional Calibration Kit available 
from Valtronics. In most cases, by carefully selecting your calibration site and time, the fresh 
air will be sufficient. The calibration site should be selected to be outdoors with fresh air. 
Avoid areas with high automobile traffic, crowds of people, and/or ventilating outlets. 
 
 
 
 
 
 

                                                        Procedure 
 

Place the meter at the chosen site, press SET to turn on the unit and note the CO

display. If it 

indicates /- 25ppm it is not necessary to go any further. The accuracy is ok and it is 
not necessary to go through the entire calibration process. If the reading is out of specification, 
press and hold both CAL/Esc and MODE to enter the calibration mode. The display will now 
show “400 ppm” and “CAL” blinking. When the blinking stops the calibration is complete. 
Wait about 10 minutes until the blinking stops and the calibration is completed. The unit will 
automatically return to Normal mode. To abort the calibration, turn off the meter at anytime. 
 
 
 
 
 
 

                                                                           

Rh Calibration

 

 
The Rh feature of the meter must be calibrated with the optional Calibration Kit supplied by 
Valtronics. This kit contains 33% and 75% Rh salt solutions and instructions for use. 
 
 
 
 
 
 
 

Summary of Contents for 2011

Page 1: ...VALTRONICS IAQ Monitor Model 2011 Instruction Manual 990116...

Page 2: ...eatures 3 Applications 3 Display 4 Key Pad 5 OPERATION 6 Power Operation 6 Measurements 6 Data Hold 6 Backlight 6 STEL TWA MIN and MAX 6 Alarm 6 Auto Power Off 7 Set up 7 CO2 Alarm 7 Temperature Scale...

Page 3: ...outstanding reliability and long term stability The unit is powered by four AA batteries or a DC adapter When the DC Adapter is used it cuts off the batteries The Adapter can not be used as a battery...

Page 4: ...TWA 8 hours of weighted averages STEL 15 minutes weighted averages Audible CO2 warning alarm Backlight on display Easy manual calibration for CO2 and humidity PC connector via RS232 interface Applica...

Page 5: ...8 hours STEL Short term exposure limit 15 minutes weighted average HOLD Readings are frozen unchanged MIN MAX Minimum maximum readings Low battery indicator DP Dew point temperature Air Air temperatu...

Page 6: ...nters CO2 calibration with MODE Enters RH calibration with DP WBT Freezes the current readings Cancels data hold function Activates or cancels the backlight Selects unit or increases value in set up S...

Page 7: ...the HOLD button will freeze all readings except the STEL and TWA Pressing the HOLD button again will cancel the hold mode Backlight Pressing and holding the MODE button for more than 1 second will ac...

Page 8: ...et the CO2 alarm note the CO2 display will be blinking at the current alarm level Each press of the MODE button will increase the value by 100 ppm Pressing the DP WBT will decrease the alarm value The...

Page 9: ...it indicates 400ppm 25ppm it is not necessary to go any further The accuracy is ok and it is not necessary to go through the entire calibration process If the reading is out of specification press and...

Page 10: ...Resolution 0 1 Rh Accuracy 3 from 10 to 90 Warm up time 30 seconds Operating temperature 0 to 50 C Storage temperature 20 to 60 C Humidity 0 to 95 Rh non condensing Power AA batteries 4X Battery life...

Page 11: ...pering of internal electronics RETURN POLICY Valtronics will honor warranty repairs within one year of purchase and will make every attempt to repair non warranty damage defects or functional failures...

Page 12: ...Copyright 2011 VTI Valtronics Inc 11 Notes For ordering information contact Valtronics Sales Department Phone 209 754 0707 FAX 209 754 0104 e mail vti val tronics com www val tronics com...

Page 13: ...Copyright 2011 VTI Valtronics Inc 12 Notes For ordering information contact Valtronics Sales Department Phone 209 754 0707 FAX 209 754 0104 e mail vti val tronics com www val tronics com...

Reviews: