![Valleylab Force 2 User Manual Download Page 31](http://html1.mh-extra.com/html/valleylab/force-2/force-2_user-manual_3246486031.webp)
Setting Up the Force 2 Generator for Surgery
Instructions for Use
Force 2 User’s Guide
4
-
3
Handswitching pencils –
connect to the
Monopolar Handswitching
or
Monopolar Accessory
receptacle.
Monopolar handswitching forceps –
connect to the
Monopolar
Handswitching
or
Monopolar Accessory
receptacle.
Power control pencils –
connect to the
Monopolar Handswitching
receptacle.
Bipolar handswitching forceps –
connect to the
Microbipolar
receptacle.
Monopolar footswitching accessories (e.g., endoscopes, forceps, or chuckhandles) –
connect to the
Monopolar Accessory
receptacle.
Bipolar footswitching forceps –
connect to the
Microbipolar
receptacle.
8. Press the
Standby
button to prevent inadvertent activation.
Dashes appear on the power setting display, and all the settings are
retained in memory.
When the surgeon is ready to use the generator:
A.
Press the
Ready
button.
B.
Press the handswitch or footswitch to activate the generator.
Warning
The
Accessory
receptacle is designed for connecting either a handswitch (three-
pin) or footswitch (one-pin) accessory, but not both at the same time. Connecting
more than one accessory to the
Accessory
receptacle will activate both
accessories simultaneously.