Section 3 | Installation 

 

 

©

 2022 

– Valeport Ltd 

Page | 6 

 

  Real Time Operation 

For real time data output, connect the signal cable to the 10 pin Subconn connector on the instru-
ment.  All Valeport signal cables include a suspension point for strain relief, and a similar arrange-
ment is recommended for other cable types.  Connect the top end of the cable to a PC using the 
appropriate method as described above. 
 

  Self-Recording Operation 

For self-recording only deployments, the instrument is switched on by insertion of the SubConn 
style switch plug.  This plug must be inserted for the unit to operate. 

the switch plug is NOT just a dummy plug; it contains links between some of the pins as de-
scribed in Section 5, which are used to turn the instrument on 

 

  LED Flashing Sequence 

The M

IDAS

 ECM is fitted with an LED visible through a polycarbonate window in the battery pack.  

The LED will flash as detailed below to indicate various states. 

Continuous ON for 15 seconds  Occurs when the Switch Plug or cable is connected, indicating 

that instrument is on and awaiting communication 

1Hz Continual Flashing 

Insufficient Power.  Change internal batteries, or completely re-
move external power, and apply higher voltage 

Continuous ON for 2 seconds 

Indicates the start of Burst Sampling pattern 

5 Rapid Flashes 

Indicates the end of Burst Sampling pattern.  The duration of the 
burst may be calculated as the time between the start and stop 
LED sequences, less 5 seconds. 

 

3.3.  Altimeter Configuration 

Systems can be specified with Tritech PA series Altimeters. For the Midas ECM to correctly com-
municate with the Altimeter, it needs to be configured as follows: 

Communication 

RS232 

Use 

Free Run ASCII/Zero No Detect (ZNE) 

Data Format 

9600, 1, 8, 1 

Power 

12V 

 

 

 

Summary of Contents for MIDAS ECM

Page 1: ...t be copied reprinted or reproduced in any material form either wholly or in part and the contents of this document and any method or technique available there from must not be disclosed to any other...

Page 2: ...tion 6 4 Recovery 7 5 Sample Battery Lifetime Calculations 8 5 1 Calculations Based on Memory 8 Continuous Data Sampling 8Hz 8 Burst Sampling 4Hz 8 Profiling 8 A Note About Removable Memory 9 5 2 Calc...

Page 3: ...rings DataLog Express Software CD Operating Manual Transit case In addition the following components may be supplied as optional extras RS485 communications adapter RS422 communications adapter FSK m...

Page 4: ...s Resolution 0 002 C 0 001m s The MIDAS ECM is technically capable of supporting other sensor types such as Turbidity Dissolved Oxygen Fluorometer Altimeter etc Such sensors require additional user at...

Page 5: ...3m Y lead 10 pin male Subconn line type MCIL10M to instru ment 2 x 4mm bunch pins to external power 9 pin female D type to PC Switching Plug 10 pin male Subconn line type MCIL10M with lock ring Note...

Page 6: ...type connector which should plug directly into a spare communications port on the back of the PC If non RS232 communications are to be used via the optional RS485 RS422 or FSK methods then the approp...

Page 7: ...ed This is so the compass module is fitted correctly It is imper ative that the instrument is deployed in the correct orientation The housing will be marked accordingly The MIDAS ECM is supplied with...

Page 8: ...ce The MIDAS ECM is fitted with an LED visible through a polycarbonate window in the battery pack The LED will flash as detailed below to indicate various states Continuous ON for 15 seconds Occurs wh...

Page 9: ...g attention to the face and electrodes of the EM sensor the core of the conductivity sensor if fitted and the face of the sound veloc ity sensor if fitted Be careful not to damage the sensors A high p...

Page 10: ...or obviously de pends on sampling scenario Here are three examples Continuous Data Sampling 8Hz Memory used per second is 6 x 8 bytes 48 bytes Total memory fitted is 16 777 216 bytes Seconds before me...

Page 11: ...ry profile to get the memory card out All memory cards are susceptible to ESD shock static electricity while being handled We take the view that the value of your data means it shouldn t be exposed to...

Page 12: ...fficient total available charge 16 6Ah Continuous data sampling 8Hz At 12V the instrument draws 140mA when sampling Total charge available is 10500mAh Number of hours available is therefore 10500mAh 1...

Page 13: ...evice Whilst this risk is small it does exist and therefore must be considered The risk could be overcome by adding an air vent to the housing but this could compromise the water tight nature of the h...

Page 14: ...ch should be checked periodically for damage and replaced if necessary 6 1 Changing Batteries The MIDAS ECM accepts 8 x D cells of either 1 5V alkaline or 3 6V Lithium type These cells are arranged in...

Page 15: ...care not to scratch the bore of the tube 6 A lead connects the battery pack to the electronics inside the tube This may be disconnected at the battery pack if required for ease 7 Replace the batterie...

Page 16: ...lternative source be sure to obtain the correct material signified by the last 4 digits of the O ring code number O ring size 200 158 4470 O Rings To help preserve the watertight nature of the equipme...

Page 17: ...NC 10 Link to Pin 1 7 2 3m Y Lead RS232 END 1 10 Way SubConn MCIL10M DLSA M END 2 BANANA PLUG END 3 9 Way D TYPE FUNCTION PIN CONNECTOR PIN PIN 1 Black 4mm Plug 1 V SUPPLY 2 Red 4mm Plug 2 V SUPPLY 3...

Page 18: ...3 9 RS422 TxA 4 10 RS422 TxB 5 11 RS422 RxA 6 12 RS422 RxB 7 2 RS232 Tx To PC 8 3 RS232 Rx From PC 9 5 5 link to 1 6 8 9 RS232 Ground 10 SHELL SHELL Internal Battery Enable 7 4 Altimeter Interface op...

Page 19: ...yment frame SubConn switch plug DataLog x2 Windows software package 3m Y lead probe to PC and power supply RS232 USB adapter tool kit operating manual and system transit case 0808008A Self recording D...

Page 20: ...ommunications 0808008B MIDAS ECM Acetal housing standard package with FSK communications 0808008C XX MIDAS ECM Acetal housing standard package with Pressure 0808008D 50 MIDAS ECM Acetal housing standa...

Page 21: ...re may actually adjust the local flow conditions so it is a good idea to fix the instrument where it will be in the prevailing current as far as possible Secondly if the structure is metallic it may h...

Reviews: