© 2020 Valeport Ltd
0660881c - fastCTD Profiler
Page 8
2.4.4. Rhodamine
Performance
Excitation:
520 nm
Detection:
650 nm
Dynamic Range:
0-1000 ppb
2 gain settings, 0-50, 0-1000
software controlled
Instrument
Detection limit:
<0.01 ppb*
Actual
Detection limit:
0.06 ppb**
Linearity:
0.99 R2
Response Time:
0.03 to 2 s
Output Rate:
0.5 Hz to 32 Hz (free running) - software controlled
* 3x SD in RO water
** Calibrated against Rhodamine solution