6. Type the reference pressure values and the corresponding corrections.
Calculate the corrections according to the following equation. You can also change the
settings on the left.
corr = P
reference
- P
measured
The values are in hPa. The reference value is as whole number and the
correction with 2 decimals.
7. To save the corrections, select
Write PTB210
.
Chapter 5 – Calibration and adjustment
23