Figure 13 Diagnostics page
4.2.1 Diagnostic values
The diagnostic values provide additional information about the measurement.
•
CALC
is the calculated concentration value without the field calibration adjustment.
•
QF
or
Quality Factor
is a value in the range 0
‑
200. The QF value depends on process
medium optical properties. A typical good value is 100, but there are process media
where 50 is acceptable. The QF value should stay constant during the process. Falling QF
value may indicate prism coating.
•
CCD
gives the position of the shadow edge on CCD in %.
•
LED
is a measure of the amount of light from the light source in %. The value should be
below 100 %.
•
BGlight
indicates the amount of outside light reaching the CCD. A value above 120
triggers a warning. A value above 240 triggers
OUTSIDE LIGHT ERROR
.
•
Traw
is the uncompensated process temperature.
4.3 Parameters
All the refractometer’s functional parameters can be changed using the
Parameters
page. The
link to the
Parameters
page is located in the menu displayed on the right- hand side of the
computer display.
PR-43 Series User Guide
M212455EN-B
18
Summary of Contents for K-PATENTS PR-43
Page 1: ...M212455EN B User Guide Process Refractometer PR 43 Series...
Page 24: ...Figure 18 Verification page PR 43 Series User Guide M212455EN B 22...
Page 50: ......
Page 52: ......
Page 53: ......
Page 54: ...www vaisala com...