USER'S GUIDE____________________________________________________________________
16 __________________________________________________________________ M210474EN-B
The following numbers refer to above:
4 = Fitting
body
NOTE
Take care not to over tighten the clasp nut to avoid difficulties when
opening it.
CAUTION
Take care not to damage the probe body. A damaged body makes the
probe head less tight and may prevent it from going through the clasp
nut.
CAUTION
In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.
Stage 1
Stage 2
1. probe
2. Tighten first manually;
probe is then sliding easily.
Finally tighten with a spanner
about 60
°
more, to have a
stable installation, Note: do not
overtighten this screw!.
3. handle of the ball valve
4. ball of the ball valve
5. process chamber / pipeline
6. manual press tool
7. the groove on the probe
indicates the upper adjustment
limit
5
2
3
4
1
6
7
3
Valve
closed