User's Guide _______________________________________________________________________
20 ___________________________________________________________________ M211399EN-G
DIP Switches of Analog Output Models
1111-066
Figure 7
DIP Switch Settings of Analog Output Models
DIP Position
Setting
1
Non-metric Non-metric units (°F).
Metric
Metric units (°C).
2
Td
Td (dewpoint) as humidity parameter on display and
analog output. Sets analog output scaling
to -20 ... +55 °C.
RH
RH (relative humidity) as humidity parameter on display
and analog output. Sets analog output scaling to
0 ... 100 %RH.
3
0...5V
0...5V analog output (both channels).
0...10V
0...10V analog output (both channels).
4
Relay On
Relay enabled.
Relay Off
Relay disabled.
5
Relay High Relay closed when measurement above setpoint.
Relay operation is linked to RH on HMW93, and T on
TMW93.
Relay Low Relay closed when measurement below setpoint.
6
Not used
7
Not used
8
Custom
Configuration through service port only.
DIP
Configuration by DIP switches only.
NOTE
DIP switch 2 does nothing on TMW92 and TMW93.
NOTE
If DIP switch 8 is set to
Custom
, the transmitter ignores all other DIP
switch settings. In custom mode the transmitter uses settings that are
configured in software using the service port. Before changing the
position of this switch, see section Changing Between DIP and Custom
No
n-M
etr
ic
Td 0..
.5V
Re
lay
On
Re
lay
Hig
h
Cu
sto
m
Me
tric RH
0..
.10
V
Re
lay
Of
f
Re
lay
Lo
w
DI
P
1 2 3 4 5 6 7 8
No
n-M
etr
ic
Td
Cu
sto
m
Me
tric RH
DI
P
1 2 3 4 5 6 7 8
HMW92/TMW92
HMW93/TMW93